Shutdown button under Utilities does not shutdown scanner and the MOD will not detach
Symptom
Shutdown button under Utilities does not shutdown scanner and the MOD will not detach with Optical disk mount/dismount errors See Long Description for more details Shutdown button under Utilities does not shutdown scanner and the MOD will not detach with Error: 124050 Optical disk system error, Error: 124048 Unable to dismount, media remains mounted and , < out of inodes > seen in /usr/adm/messages Customer tried to do their daily system shutdown by going into Utilities and then pressing shutdown. When they pressed the button the screen flashed, but the system does not Shutdown. They were also unable to detach the MOD disk. When they pressed L1-b they observed the following scrolling messages Oct 4 11:20:00 CT01_OC0 vmunix: /: out of inodes Oct 4 11:21:00 CT01_OC0 vmunix: /: out of inodes Oct 4 11:21 /: out of inodes Oct 4 11:23:01 CT01_OC0 vmunix: /: out of inodes Oct 4 11:26:00 CT01_OC0 vmunix: /: out of inodes Oct 4 11:29:00 CT01_OC0 vmunix: /: out of inodes Typing sd in the text window did not shut the system down. The following message was displayed genesis @ CT01_OC0 1: sd /: create/symlink failed, no inodes free /usr/g/bin/ShutDownGenesis: /tmp/sh115780: No space left on device genesis @ CT01_OC0 2:
Error message
Error: 124048 Unable to dismount, media remains mounted. See LD for complete error Error: 124048 Unable to dismount, media remains mounted. Mon Oct 6 15:50:12 1997 Suite: CT01 Host: OC0 Proc: odc0 Error: 124048 File: ODController.m 1.13 Method: -dismount Line: 2319 Unable to dismount, media remains mounted. err: 124048 errno: 48 don_errno: 0 out of inodes seen in /usr/adm/messages See LD for complete error < out of inodes > seen in /usr/adm/messages out of inodes message can also be seen scrolling in the text window when hitting L1-b Oct 4 11:20:00 CT01_OC0 vmunix: /: out of inodes Oct 4 11:21:00 CT01_OC0 vmunix: /: out of inodes Oct 4 11:21 /: out of inodes Oct 4 11:23:01 CT01_OC0 vmunix: /: out of inodes Oct 4 11:26:00 CT01_OC0 vmunix: /: out of inodes Oct 4 11:29:00 CT01_OC0 vmunix: /: out of inodes Error: 124050 Optical disk system error. See LD for complete error Error: 124050 Optical disk system error. Mon Oct 6 15:50:01 1997 Suite: CT01 Host: OC0 Proc: odc0 Error: 124050 File: ODController.m 1.13 Method: rmvMountEntries() Line: 2135 Optical disk system error. err: 28 errno: 28 don_errno: 0
Applies to
- Product
- HSA Z Series HSA Z Series HSA Z Series Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-301976P1 Seagate Wren 5 Disk Drive 600MB See LD for details 46-301976P1 Seagate Wren 5 Disk Drive 600MB Part #: 46-301976P1 Part Type: GP REPAIRABLES Part Weight: 5.00 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: SEAGATE WREN 5 DISC DRIVE, 600MB FRU: Yes Status: Orderable Parts Offered as: Current: 46-301976P1 Harvested: Not Available Exchange: Not Available Comments: FLS - FE: PLEASE REFER TO SN T-1326 REGARDING GENESIS COMPUTER; CONSOLE TROUBLESHOOTING TIPSCT SPUR TEAM 9/94 PER GREG MAHALICK THERE IS NO FIRMWARE ON THIS DRIVE 9/12/97 DESC: 600 MB, WREN 5, DISK DRV REPAIR OF T3202AM OUTSOURCED FROM DBR TO ICS.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.