Pular para o conteúdo principal
GE Healthcare · CTLightSpeed1822152730917090018486215006182211

LightSpeed 2.X (Plus, QX/i) - Octane

Symptom

Cradle Drive Motion was prohibited by hardware Interference switches

Error message

Error: 182215 Motion was prohibited by hardware interferenceHardware Inhibit Switches 0x80fc. Mon Sep 27 20:29:03 2004Error: 182215 Host: ETC Process: Cradle File: cmcStart.c 1.60 7/23/96 7/23/96 Line: 404 Function: PATIENT ANATOMY POSITIONING : Cradle Drive Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 20:29:03:355 Log Series: 32 HW/SW error: Motion was prohibited by hardware Interference switches. Hardware Inhibit Switches 0x80fc Current Position 27309 (.01 mm) Dir: 1 (0 = Out 1 = In [Gantry]) Positions 0 (.01 mm) 170900 (.01 mm) Mon Sep 27 20:29:03 2004 Error: 182215 Host: ETC Process: Cradle File: cmcStart.c 1.60 7/23/96 Line: 404 Function: PATIENT ANATOMY POSITIONING : Cradle Drive Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 20:29:03:355 Log Series: 32 HW/SW error: Motion was prohibited by hardware Interference switches. Hardware Inhibit Switches 0x80fc Current Position 27309 (.01 mm) Dir: 1 (0 = Out 1 = In [Gantry]) Interference Positions 0 (.01 mm) 170900 (.01 mm)

Error: 184862 Cradle unable to initiate a profile. Desired position 3880, velocity 47 internal cradle status 15006 See Long Description for full error Error: 184862 Host: ETC Pr

Applies to

Models
LightSpeed
Product
LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2269177 Cradle latch/lock assembly FOR CATHAY-TABLE 2269177 Cradle latch/lock assembly See LD for picture of location 2269177 Cradle latch/lock assembly Part #: 2269177 Part Type: GP CONSUMABLES Part Weight: 2.20 LBR Repairable: No Returnable: No Hazardous Material: N Description: LATCH ASSEMBLY FOR CATHAY-TABLE FRU: Yes Status: Orderable Parts Offered as: Current: 2269177 Used: Not Available Exchange: Not Available Comments: FLS - LATCH ASM ADDED. USED ON H2 TABLE LCC. ASIA HAS IDENTIFIED THIS AS A LINKED PART (SEE 46-296233G2).AWAITING CONFIRMATION ON GL18.MSA8429 11/8/01 START DEPLOYMENT FROM ASIA ON FW7, 02. MS20544 1/31/02
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.