LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

The Scanner is fine during the monitor phase of smart prep but fails during Scan Phase See Long Description: The Scanner is fine during the monitor phase of smart prep but fails during Scan Phase The Scanner proceeds through the monitor phase of Smart Prep and then tranistions to the scan Phase, the Green Scan button on the SCIM Illuminates, then the technologist presses the green scan button and the scan hardware aborts the scan.

Error message

Error: 245231 Error: smartPrep transition to scan phase failed: Message: begin_scan() failed for first diag scan - See Long Description: Wed Jan 25 15:26:13 2006 Error: 245231 ct01 scanmgr scanmgr_lib.cxx 3268 Error: smartPrep transition to scan phase failed: Message: begin_scan() failed for first diag scan This about sums it up: Error: smartPrep transition to scan phase failed: Look for additional detail in the gesyslog.

Error: 245229 Transition out of Monitor phase successfully. Starting first diag scan with AUTOSCAN ON. See Long Description for complete error. Wed Jan 25 15:26:12 2006 Error: 245229 ct01 scanmgr scanmgr_lib.cxx 3240 Info: Transition out of Monitor phase successfully. Starting first diag scan with AUTOSCAN ON. Exam: 4026 Monitor Series: 200 Diagnostic Series: 3.

Error: 260118320 E_EBL_IN line is disable. IO port D register bitmap. b8 See Long Description: Wed Jan 25 15:26:12 2006 Error: 260118320 orp orp_subsys gen_exec.c 3895 Wed Jan 25 15:26:10 2006 Host : On-Board Rotating Processor Ermes # : 260118320 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 3895 Function : No System Function Reported Scan Type : Axial Scan: 4026/200/14 E_EBL_IN l

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Software
05MW31.6 for LightSpeed VCT 7.X with Xtream FX GRE-GOC5 Console Introduced with LS VCT M3A systems produced from Oct 1st, 2005 05MW31.6 VCT 3 disk set 5149029 Includes: - Operating System Disk 5134309 - Application software Disk 5149024 - Serial Over LAN 2399297 Part Number for reference: 31.6 Class C 5149025 31.6 Class M 5149026 5149029 05MW31.6 VCT M3A Collector 3 disk set includes: 5134309 Operating System SW 5149024 Application SW 2399297 SOL Serial Over Lan SW 5149025 05MW31C.6 Class C M3A Proprietary SW 5149026 05MW31M.6 Class M M3A Proprietary SW See list of standard Options
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues