GE Healthcare · CTLightSpeed244609260007774260007768
LightSpeed 4.X - GRE
Symptom
Scanner Hardware Stopped Scan message during axial scan. Long Description for Feedback Only. Scanner Hardware Stopped Scan message during axial scan
Error message
Error: 244609 WARNING - Invalid DAS is detected. The gantry can be down or the DAS can be bad See LD for complete error Mon Jan 23 07:02:57 2006 Error: 244609 Host: wct1 Process: fwmgr File: HwSetUp.c Line: 4117 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 WARNING - Invalid DAS is detected. The gantry can be down or the DAS can be bad Error: 260007774 List of Converter Cards From the Previous Error Message Card List: xx xx See Long Description for full error Tue Dec 30 09:25:17 2008 Error: 260007774 Host: orp Process: orp_subsys File: CnvBoardInterface.cc Line: 852 Tue Dec 30 09:25:17 2008 Host : Das Ermes # : 260007774 Exception Class : Secondary Severity : Sec/Soft File : CnvBoardInterface.cc Line# : 852 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 List of Converter Cards From the Previous Error Message Card List: 8,11,14-15,19,21,25-27,29-36,38-58,60-63 Note on this error. Sometimes the actual bad board will be the one that is not listed. Error: 260007768 DAS Converter Boards are not installed in valid configuration See LD for complete error Mon Jan 23 07:02:45 2006 Error: 260007768 Host: DAS Process: D
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-170021P81 20A M-DAS power supply fuse See Long Description for details Part #: 46-170021P81 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: LITTELFUSE 323020, 20A,250V, 3AB FRU: Yes Status: Orderable Comments: FLS - 20A,250V
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.