Pular para o conteúdo principal
GE Healthcare · CTLightSpeed26000650050038

LightSpeed 4.X - GOC2

Symptom

Unable to scan

Error message

Error: 260006500 Axial AC Drive reported fault code: F60 - Encoder loss. Encoder Loss indicates that the axial drive can not sense the axial encoder See long description for full error Sun Jan 08 11:00:27 2006 Error: 260006500 Host: STC Process: Axial File: AX2fault.c Line: 196 Function: AXIAL CONTROL : axial control Scan: 50038/8/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 11:00:26:535 Log Series: 35 The axial AC Drive reported fault code: F60 - Encoder Loss This indicates that the axial drive can not sense the axial encoder. Possible causes are: Encoder, cabling, unplugged axial control board or improper axial drive encoder jumper setting.

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GOC2 405L.8_H4.2M4 LightSpeed 4.X with GOC2 and MDAS 16 Gantry H4 Gantry Model # 2339985 and Global Operator Console 2 Model #'s 2366438 and 2366438-2(Hino) with HP8000 CPU (or newer) and Pegasus Recon ICE Box Product: LightSpeed16 405L.8_H4.2M4 CPU: GenuineIntel Intel(R) Xeon(TM) CPU 2.66GHz 4 2657 MHz i686 Processors Main memory size: 2048 Mbytes [root@presbyct2]# sg_map -i /dev/sg0 /dev/sda SEAGATE ST336753LW HPS2 /dev/sg1 /dev/sdb SEAGATE ST373453LW HPS1 /dev/sg2 /dev/sdc SONY SMO-F551-SD 1.05 /dev/sg3 /dev/scd0 MATSHITA DVD-RAM SW-9574S AZX8 /dev/sg4 /dev/scd1 HL-DT-ST DVD-ROM GDR8163B 0B15 [root@presbyct2]# df -h Filesystem Size Used Avail Use% Mounted on /dev/sda2 4.0G 1.7G 2.3G 41% / /dev/sda1 27M 12M 14M 44% /boot none 1008M 0 1007M 0% /dev/shm /dev/sda6 14G 3.5G 10G 25% /usr/g /dev/sda7 9.8G 753M 9.0G 8% /usr/g/sdc_image_pool /dev/sdb1 68G 28G 40G 41% /usr/g/sdc_image_pool2 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2236254 Axial Control Board - used in H2 gantry STC 2236254 Axial Control Board 2236254 Axial Control Board Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: AXIAL CONTROL BOARD FRU: Yes Status: Orderable Comments: FLS - H2 GANTRY: COMMON TO MFG MS26561 DSM 1/7/01 COLLECTION OF DEFECTIVES START7/23/01 MS5277 FLS - PLEASE CONVERT SPH ORDER TO OVERNIGHT SHIPMENT IF PART AVAILABLE IN MEMPHIS AND IF FEDEX SERVICE (INCLUDING 8:00 AM SERVICE) MEETS FE NEED TIME CRITERIA
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.