Service Note 43541_2 Jan04 All Elscint Ring Gantry Actuators
Symptom
Service Note 43541_2 Jan04 All Elscint Ring Gantry Actuators: Actuator Life Expectancy Verification Procedure Service Note 43541 _2 Jan04 All Elscint Ring Gantry Actuators: Actuator Life Expectancy Verification Procedure Problem Description A recent incident was reported where unintentional movement of a detector almost caused injury to a patient. The actuator was sent to the factory for inspection and it was found that the failure occurred due to natural wear to an internal nut which has a life expectancy of about 10 years. Notes:
Applies to
- Product
- APEX SPX APEX SPX Gamma Camera (Elscint) APEX SPX Gamma Camera (Elscint)
- Component
- *Gantry Gantry Gantry
- Area
- Nuclear Medicine
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare Nuclear Medicine issue are available to FSELIB subscribers — re-authored, web-first, searchable.