Exciter Calibration Errors And Power Monitor Trips Due to CERD Exciter

Symptom

Exciter Calibration Errors And Power Monitor Trips Intermittent exciter calibration errors and 85 faults, FE performed power monitor functional checks and they passed.

Error message

2234496 No or little signal amplitude detected during Exciter calibration. Fri Jan 30 16:21:12 1998 Error: 2234496 Host: oc001 Process: NSP File: TPS:cie_exc_cal Line: 455 No or little signal amplitude detected during Exciter calibration. Possible hardware problem - please run CERD Diagnostics!

2224711No successful FTG check in 1 iterations Wed Sep 23 12:07:20 1998 Error: 2224711 Host: MRS1 Process: NSP File: TPS:aps_exec Line: 10 No successful FTG check in 1 iterations (hold_aps_ta: 120 final TA: 64).

2225664 RF Amplifier fault 85 RF Transmit power has reached safety limit allowed for scan parameters and patient weight Error: 2225664 RF Amplifier fault 85. RF Transmit power has reached safety limit allowed for scan parameters and patient weight. Check patient weight entered. amplitude: 15 duty cycle: 05 Fault in head amplitude. If you need to confirm the source of the fault, reset the power monitor and repeat the scan.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
8.4
Area
MR