Indigo Computer Fails to Boot up.
Symptom
Indigo Computer Fails to Boot up. The LED on the Computer is OFF. No Video is seen on the Monitor see long description LX system with an Indigo computer running 8.3 M5 Software. The Indigo computer fails to boot up. The led on the front of the computer is OFF. The fan is running and the led is lit on each of the SCSI peripherals, the CD, The floppy, and the MOD. No video is seen on the monitor. Checked the Aux power supply voltages and all are good. The documented procedure to check the SGI power supply is by verifying power on diags pass. The led on the front of the computer is OFF, so it appears power on diags failed. Indigo troubleshooting shows the following for power on Diagnostics with unlit led: Unlit led, no error - possible power supply failure, possible CPU failure.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 8.3 8.3
- Component
- 2115457-28 BOARD SET IP22–GEMS W/250MHZ R4400 CPU MOD & AUDIO for Indigo computer. See long description ** FE ASSIST ** 1. You must push the power reset switch located on the Paddle Board or the system will not boot when the Console is powered on. ** 2. The IP22 is tested with a known good set of Simm memory modules. New memory modules must be ordered separately. **16 MB Simm Memory Part Number: 2115457-22. **32 MB Simm Memory Part Number: 2115457-23. **3. For assistance on T/S CPU boot problems refer to CTi Service Tips Direction 2160288-100 Minimum Boot Configuration.(212001261 7/11/07)
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.