Pular para o conteúdo principal
GE Healthcare · MRSigna22415182241614224150822415442241510

Will Not Scan with Gradient Errors

Symptom

Will Not Scan with Gradient Errors.

Error message

2241518 The GIP reported a GPM Power Supply Fault. 2241518 The GIP reported a GPM Power Supply Fault.File: HPC_PORT_ID:GAP/GIP/GP Line: 2359 The GIP reported a GPM Power Supply Fault. Check the GPM or the GPM Power Supply.->

2241614 GPM Power Supply V_LO Monitoring Error. Mon Jun 25 23:45:31 2001 Error: 2241614 Host: lx-mr Process: NSP File: HPC_PORT_ID:GAP/GIP Line: 546 GPM Power Supply V_LO Monitoring Error. Check GRAM Tuning, GPM Power Supply, incoming power, & GIP. Low: 100.000V High: 220.000V Value: 94.352V Axis: N Standby/Ready State: S

2241508 The GIP is reporting ? Axis GPM Power Off Wed Jun 30 07:42:27 2004 Error: 2241508 Host: SJHMR Process: NSP File: HPC_PORT_ID:GAP/GIP/GP Line: 303 The GIP is reporting X Axis GPM Power Off. If all 3 axes are reporting this fault, check the GPM Power Supply circuit breakers and cables. Check GPM fuses and the cable between the GIP and the GPM.

2241544 The gradient system on the X axis recovered successfully. Mon Jun 25 23:45:30 2001 Error: 2241544 Host: lx-mr Process: NSP File: HPC_PORT_ID:GAP/GIP Line: 1342 The gradient system on the X axis recovered successfully. There are no faults at the present time.

2241544 The gradient system 

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Component
2154416-3 GRADIENT POWER MODULE POWER SUPPLY
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.