GE Healthcare · MRSigna22415182241614224150822415442241510
Will Not Scan with Gradient Errors
Symptom
Will Not Scan with Gradient Errors.
Error message
2241518 The GIP reported a GPM Power Supply Fault. 2241518 The GIP reported a GPM Power Supply Fault.File: HPC_PORT_ID:GAP/GIP/GP Line: 2359 The GIP reported a GPM Power Supply Fault. Check the GPM or the GPM Power Supply.-> 2241614 GPM Power Supply V_LO Monitoring Error. Mon Jun 25 23:45:31 2001 Error: 2241614 Host: lx-mr Process: NSP File: HPC_PORT_ID:GAP/GIP Line: 546 GPM Power Supply V_LO Monitoring Error. Check GRAM Tuning, GPM Power Supply, incoming power, & GIP. Low: 100.000V High: 220.000V Value: 94.352V Axis: N Standby/Ready State: S 2241508 The GIP is reporting ? Axis GPM Power Off Wed Jun 30 07:42:27 2004 Error: 2241508 Host: SJHMR Process: NSP File: HPC_PORT_ID:GAP/GIP/GP Line: 303 The GIP is reporting X Axis GPM Power Off. If all 3 axes are reporting this fault, check the GPM Power Supply circuit breakers and cables. Check GPM fuses and the cable between the GIP and the GPM. 2241544 The gradient system on the X axis recovered successfully. Mon Jun 25 23:45:30 2001 Error: 2241544 Host: lx-mr Process: NSP File: HPC_PORT_ID:GAP/GIP Line: 1342 The gradient system on the X axis recovered successfully. There are no faults at the present time. 2241544 The gradient system
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Component
- 2154416-3 GRADIENT POWER MODULE POWER SUPPLY
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.