GE Healthcare · MRSigna2241516

All three miniGRAMs have Undervoltage Faults and fail to go to ready when attempting to scan.

Clarify #011192 5X,LX

Symptom

All three miniGRAMs have Undervoltage Faults and fail to go to ready when attempting to scan. System is an LX with High Slew SGD gradients.

Error message

2241516 "The GIP reported a Z Axis miniGRAM UnderVoltage Fault 2241516 "The GIP reported a Z Axis miniGRAM UnderVoltage Fault. Try swapping the miniGRAM Control Board, or the entire miniGRAM axis, or the miniGRAM Power Supply to isolate the problem

2241516 "The GIP reported a X Axis miniGRAM UnderVoltage Fault. Try swapping the miniGRAM Control Board, or the entire miniGRAM axis, or the miniGRAM Power Supply to isolate the problem

2241516 "The GIP reported a Y Axis miniGRAM UnderVoltage Fault. Try swapping the miniGRAM Control Board, or the entire miniGRAM axis, or the miniGRAM Power Supply to isolate the problem

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR