GE Healthcare · CTLightSpeed26000151450004260001007260001022260001012
LightSpeed 4.X - GRE
Symptom
Scan hardware stopped scan with RCIB errors
Error message
Error: 260001514 RCIB Bus: Fault Line Interrupt Occurred. Timeout Waiting for Host Response to Fault Line Verification. See long description for full error Wed Apr 27 01:00:51 2005 Error: 260001514 Host: orp Process: orp_subsys File: notiftsk.c Line: 1368 Running: System Reset Wed Apr 27 01:00:51 2005 Host : Das Ermes # : 260001514 Exception Class : Abort Severity : Secondary File : notiftsk.c Line# : 1368 Function : No System Function Reported Scan Type : None/Unknown Scan: 50004/1/0 RCIB Bus: Fault Line Interrupt Occurred. Timeout Waiting for Host Response to Fault Line Verification. (Timeout:500 msecs). Error: 260001007 CAN serial link error. A CC error was detected when sending a message to the CCB See the long description for the full error code Tue Nov 13 21:06:38 2007 Error: 260001007 Host: OBC Process: HOP File: hop_cc_send.c Line: 196 Function: No System Function Reported Scan: 2552/1/0 Type: None/Unknown Exception Level: Pri/Soft Time: 21:06:38:745 Log Series: 937 CAN serial link error. An CC error was detected when sending a message to the CCB. Message code: 1004 Node ID: 52 Object ID: 0 Send status: 5 Error: 260001022 Communications between the DCB sub-system and OBC
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2179860-2 HEMRC CONTROL BOARD See LD for more info 2179860-2 HEMRC CONTROL BOARD. Part #: 2179860-2 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: HEMRC CONTROL BOARD FRU: Yes Status: Orderable FEMC PART COMMENTS: FLS - HEMRC CONTROL BOARD. REPLACES P/N 2179860. ADDED PER PCN 220717.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required