Unable to scan, Erbtec errors.

Symptom

Unable to scan with Erbtec errors

Error message

2225665 ERBTEC RF Amplifier Fault 86. +6kV Arc Detected Mon Oct 4 15:21:22 2004 Error: 2225665 Host: SNIMR1 Process: NSP File: HPC_PORT_ID:spi_rf_mon Line: 3060 ERBTEC RF Amplifier fault 86. +6kV arc detected. Possible defective PA tube or bad processor board.

2225643 ERBTEC RF Amplifier fault 57 +3kV power supply unacceptable within 0.5 to 6 seconds after closing HV start relay 2225643 ERBTEC RF Amplifier fault 57. +3kV power supply unacceptable within 0.5 to 6 seconds after closing HV start relay. Possible defective HV rectifier/filter board, bad relay in AC Switching Module, or bad A/D on Processor Board.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Component
46-305000P57 Erbtec Doorknob Capacitor
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.