GE Healthcare · CTLightSpeedVCT26011631950132
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
When pressing Alignment Light Button, Gantry rotates but no lasers come on and the ORP reset indicator flashes on the gantry display
Error message
Error: 260116319 State Response Timeout Occurred See Long Description for full error Fri Sep 2 07:22:58 2005 Error: 260116319 Host: tgp Process: tgp_subsys File: subsys_mgr.c Line: 2343 Fri Sep 2 07:22:55 2005 Host : Table Gantry Processor Ermes # : 260116319 Exception Class : Secondary Severity : Pri/Soft File : subsys_mgr.c Line# : 2343 Function : No System Function Reported Scan Type : None/Unknown Scan: 50132/1/0 State Response Timeout Occurred Subsystem: Table Gantry Processor State: 12 Non-Responding App Id: 100 Timeout Value: 1000 msec.
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 2105617-5 (replace 2105617-4) SINGLE, TERMINATED LASER DIODE ASM - used on H2 and HP gantry See long description for details SINGLE, TERMINATED LASER DIODE ASSEMBLY Part #: 2105617-5 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: SINGLE, TERMINATED LASER DIODE AMP CONNECTOR, 105MM LONG, PHOTONIC FRU: Yes Status: Orderable Part #: 2105617-4 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: SINGLE, TERMINATED LASER DIODE AMP CONNECTOR, 105MM LONG FRU: Yes Status: Orderable Comments: FLS - H2 GANTRY. COMMON TO MFG MS26561 DSM 1/7/01 PART OF 2229971
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.