FMI 40775_0 Preventing APEX ARG Actuator from Loosening
Symptom
FMI 40775_0 Preventing APEX ARG Actuator from Loosening FMI 40775 _0 Preventing APEX ARG Actuator from Loosening - Updates and Corrections Over a period of time, the locking nut securing the rear actuator clamp to the actuator rod vibrates loose. This could result in the rear clamp releasing from the actuator. Securing the rear actuator clamp and locknut with Loctite 262 prevents the clamp and locknut from loosening No Updates
Applies to
- Product
- APEX SPX APEX SPX Gamma Camera (Elscint) APEX SPX Gamma Camera (Elscint) APEX SP APEX SP Gamma Camera (Elscint) APEX SP Gamma Camera (Elscint)
- Component
- *Gantry Gantry Gantry
- Area
- Nuclear Medicine
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare Nuclear Medicine issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required