How to construct and apply a Pinch Sensor defeater plug
Symptom
How to construct and apply a Pinch Sensor defeater plug There's no part number to order a pinch sensor defeater plug, you have to build your own if you want one. It's not easy to find the directions and parts list to build and use a defeater plug for troubleshooting pinch sensor problems on HFO magnet enclosures. There's an error in the Magnet I/F Troubleshooting docs that describe the process. In one place 5 ohm resistors are specified and in another, 5k ohm. Actually, only 5k ohm resistors are used. PQR 13058327 & 13058328 submitted.
Applies to
- Product
- Open Speed with EXCITE For HFO4 and HFO5 only
- Software
- OPENMR4-HFO4 Software version for HFO4 system, which is OpenSpeed with Excite Technology
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- How to construct and apply a Pinch Sensor defeater plug
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- OrthoMRI Network cable is unplugged
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- CWA ID100060 Unable To Film.
- Fails to Scan with TORSO Coil.