Information on ACGD Cabinet Air Flow for proper SGA Operation.

Symptom

Information on ACGD Cabinet Air Flow for proper SGA Operation. ACGD gradient amplifiers require a minimum amount of air flow to protect the power electronics from overheating.We have found situations where the fan becomes misaligned relative to the inlet ring during shipping. Furthermore, we have found that at some sites, the cabinet configuration is out of spec (missing panels, open doors, etc.). This SMILe provides a checklist to make sure that the cabinet airflow is in spec. (SMILe Nov. 10, 2005)

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution