Pular para o conteúdo principal
GE Healthcare · CTHiSpeedLightSpeed2005015472000018212005015321500120050153515003

HiSpeed QX/i - Octane

Symptom

Smart Prep fails when monitor phase is initiated

Error message

Error: 200501547 Process: scanRx A Communication Error has occurred in uifIpcSendRecvSmartPrepStartUp Wed Dec 7 13:38:30 2005 Error: 200501547 Host: sect092 Process: scanRx File: uifSmartPrepIf.c Line: 187 Function: Data Acquisition : OC Processing A Communication Error has occurred in uifIpcSendRecvSmartPrepStartUp with seq_num = 1200 and status = 200001821

Error: 200501532 Process: scanRx timed out while attemting to receive a message Wed Dec 7 13:38:30 2005 Error: 200501532 Host: sect092 Process: scanRx File: uifIpcCommonFunctions.c Line: 2505 Function: No system function reported Service uif_scan_rx failed to receive message with event code 15001 from service uif_exam_rx_display_mgr with status: Timed out while attempting to receive a message.

Error: 200501535 Process: scanRx Service uif_scan_rx received unknown message with event code 15003 from service uif_smartpreprx See Long Description for details Wed Dec 7 13:38:30 2005 Error: 200501535 Host: sect092 Process: scanRx File: uifIpcCommonFunctions.c Line: 737 Function: No system function reported Service uif_scan_rx received unknown message with event code 15003 from service uif_smartpreprx

Error: 200107023 A Failure had b

Applies to

Models
HiSpeed, LightSpeed
Product
HiSpeed QX/i - Octane Always MDAS4 101I.2_HSQXIM5 101.4_HSQXIM4 HiSpeed QX/i - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 101I.2_HSQXIM5 101.4_HSQXIM4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
101I.2_HSQXIM5 See LD for ClassM & C 101I.2_HSQXIM5 used on HiSpeed (YMS) only Class M software - 2395173 or 2401759 (6 pack) Class C software - 2395172 or 2401759 (6 pack) Patch G.6 Mandatory FMI# 25363 (P/N 5136941)
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.