GE Healthcare · Nuclear Medicine43480
Service Note 43480_0 Oct/04 Warning Sticker on Scintibed
Symptom
Service Note 43480_0 Oct/04 Warning Sticker on Scintibed Service Note 43480 _0 Oct/04 Warning Sticker on Scintibed - Updates and Corrections ‘Weight limit’ warning stickers placed on the Scintibed carbon fiber table can interfere with proximity detectors of the wholebody autocontouring device and cause an unnecessary increase of the gantry radius due to reflection. No Updates
Applies to
- Product
- SMV DS Camera SMV DS Camera SMV DS Camera
- Component
- *Patient Table *Patient Table *Patient Table
- Area
- Nuclear Medicine