Pular para o conteúdo principal
GE Healthcare · Nuclear Medicine43520

Service Note 43520_1 Jul/09 Incompatible Statif Boards DST, DST-XL, and i-series SMV cameras

Symptom

Service Note 43520_1 Jul/09 Incompatible Statif Boards DST, DST-XL, and i-series SMV cameras Service Note 43520 _1 Jul/09 Incompatible Statif Boards In a system with FEA, the CAN board requires 11 volts. The Statif 2 supplies this voltage with a 15 volt module which is factory adjusted for 11 volts. In a system with GNCC, the CAN II board requires only 5 volts. A new Statif 2 was introduced with GNCC and is called "Statif 2 GNCC". Statif 2 GNCC has a 5 volt module. No Updates

Applies to

Product
SMV DS Camera SMV DS Camera SMV DS Camera
Component
*Gantry Gantry Gantry
Area
Nuclear Medicine

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare Nuclear Medicine issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare Nuclear Medicine issues