Service Note 43520_1 Jul/09 Incompatible Statif Boards DST, DST-XL, and i-series SMV cameras
Symptom
Service Note 43520_1 Jul/09 Incompatible Statif Boards DST, DST-XL, and i-series SMV cameras Service Note 43520 _1 Jul/09 Incompatible Statif Boards In a system with FEA, the CAN board requires 11 volts. The Statif 2 supplies this voltage with a 15 volt module which is factory adjusted for 11 volts. In a system with GNCC, the CAN II board requires only 5 volts. A new Statif 2 was introduced with GNCC and is called "Statif 2 GNCC". Statif 2 GNCC has a 5 volt module. No Updates
Applies to
- Product
- SMV DS Camera SMV DS Camera SMV DS Camera
- Component
- *Gantry Gantry Gantry
- Area
- Nuclear Medicine
Full cause & step-by-step resolution