GE Healthcare · CTLightSpeedVCT260134804
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Table will not move in or up. See Long Description for Details. Table will not move in or up with Error: 260134804 IMS tape switch OUT (front) is activated and one or more Error: 260134713, 260134714, 260134715, 260134716, 260134717, 260134718, 260134719, 260134720, 260134803, 260134804, 260134805, 260134806
Error message
Error: 260134804 IMS tape switch OUT (front) is activated see long description for complete error Wed Nov 9 16:44:37 2005 Error: 260134804 Host: tgp Process: tgp_subsys File: gtcb_imc_isrfunc.cxx Line: 116 Wed Nov 9 16:44:38 2005 Host : Table Sub-System Ermes # : 260134804 Exception Class : Secondary Severity : Pri/Soft File : gtcb_imc_isrfunc.cxx Line# : 116 Function : No System Function Reported Scan Type : None/Unknown Scan: 405/2/0 IMS tape switch OUT (front) is activated. Position is 0 um.
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5133730 Tape Switch (100mm) used as IMS tape switch OUT (front) see long description for tape sensor part numbers Part #: 5133730 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: TAPE SWITCH 1 Positioning GT FRU: Yes Status: Orderable Comments: Part add per ECR/ECO2031799 305014118 08/28/2006. source pole/planner code revised per to Global Table transfer Location Tape Switch Part Number Description Switch Description in gesyslog TOP COVER FRONT ASSY 5133731 TAPE SWITCH 2 (185mm) Elevation touch switch1 (tape switches on underside of top front cover) 5133732 TAPE SWITCH 3 (265mm) Elevation touch switch1 (tape switches on underside of top front cover) IMS LEFT SIDE COVER 5133730 TAPE SWITCH 1 (100mm) IMS tape switch OUT (front) 5133732 TAPE SWITCH 3 (265mm) Elevation touch switch4 (tape sw on underside of IMS side cover) IMS REAR UPPER COVER 5164938 TAPE SWITCH 4 (208mm) IMS Tape Switch IN� (rear) IMS RIGHT SIDE COVER 5133730 TAPE SWITCH 1 (100mm) IMS tape switch OUT (front) 5133732 TAPE SWITCH 3 (265mm) Elevation touch switch4 (tape sw on underside of IMS side cover)
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.