Symptom

Scanner Hardware Stopped Scan message DURING Scan data acquisition. Long Description for Feedback Only. Scanner Hardware Stopped Scan message DURING Scan data acquisition.

Error message

Error: 245190 System aborted Axial scanning. No views, either offset or data, were collected/received. Data views expected: 992 See the Long Description for the full error code. Thu Jan 13 06:58:23 2005 Error: 245190 Host: ct02 Process: scanmgr File: scan_complete.cxx Line: 303 Function: System Control : Scan Control Scan Type: Axial Scan: 0/0/0 Scan Seq Id: 1000 System aborted Axial scanning. No views, either offset or data, were collected/received. Data views expected: 992

Error: 183177 Abnormal Xray termination. External x-ray interlock opened See the long description for the full error code Tue Sep 6 09:47:22 2005 Error: 183177 Host: OBC Process: Generator File: gen_xray.c 1.31 Line: 604 Function: XRAY GENERATION : Xray Exposure Interlocks Scan: 637/7/1 Type: Helical Exception Level: Pri/Most Severe Time: 09:47:22:260 Log Series: 487 Abnormal Xray termination. External x-ray interlock opened.

Error: 245180 System aborted Xray-on Static scanning. Wed May 12 13:54:02 2004 Error: 245180 Host: ctpro100 Process: scanmgr File: scan_complete.cxx Line: 313 Function: System Control : Scan Control Scan Type: Xray-on Static Scan: 0/0/0 Scan Seq Id: 1000 System aborted Xray-on Static sca

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 1.X - GRE H1_P_S4 Lightspeed 1.X - GRE Method of visual identification Large square gantry similar in appearance to a CTi gantry. Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2177065 P9237HD RF TRANSMITTER ELECTRONICS See the LD for part jpg 2177065 P9237HD RF TRANSMITTER ELECTRONICS Part #: 2177065 Part Type: GP CONSUMABLES Part Weight: 0.11 LBR Repairable: No Returnable: No Hazardous Material: N Description: P9237HD RF TRANSMITTER ELECTRONICS FRU: Yes Status: Orderable P9237HD RF TRANSMITTER ELECTRONICS U/O YMS LXI,DXI,FXI & HELIOS ASIA SOURCE POLE. AM BUYS LOCALLY. HELIOS M3 4TH QTR
Area
CT

Full cause & step-by-step resolution