FMI 40719 Genie Acq Versions 2.5 and 2.5.1 Software Patches GA250-032-2 and GA250-033-0 Installation - Updates and Corre…
FMI 40719 Genie Acq Versions 2.5 and 2.5.1 Software Patches GA250-032-2 and GA250-033-0 Installation - Updates and Corrections
Symptom
FMI 40719 Genie Acq Versions 2.5 and 2.5.1 Software Patches GA250-032-2 and GA250-033-0 Installation - Updates and Corrections Update 1. After loading GA250-033-0 patch, MG Roll drive noise when roll is not moving. Making a very loud squealing or growling noise at times FMI 40719 Genie Acq Versions 2.5 and 2.5.1 Software Patches GA250-032-2 and GA250-033-0 Installation - Updates and Corrections Background It has been reported that the response time when using the Hand Held Controller (HHC) is very slow. In some cases the HHC must be pressed several times before the system starts to move. It is sometimes necessary to reboot the computer in order for HHC to work. Patch GA250-032-2 implements download of the DSP code to the ACS controller. It has been reported that the movement of one of the detectors does not stop immediately whenthe Detector Radius button is released. Patch GA250-032-2 fixes this problem. Patch GA250-033-0 fixes the radial and roll parameters for MG. Update 1 . After loading GA250-033-0 patch, MG Roll drive noise when roll is not moving. Making a very loud squealing or growling noise at times
Applies to
- Models
- Millennium
- Product
- MG/MC/Myosight Millennium MG/MC Millennium MyoSIGHT
- Software
- Genieacq 2.5 Genieacq 2.5 build 15 Software Revision: Genieacq 2.5 Build 15 Kit Part Number: FMI # / Service Note #: FMI 40687 Comments: For MG, MC, MPR, MPS, IBACQ (400A), OptimaNX. Not for MPT, IBx (X detector), or MyoSight.
- Component
- *Software Software Software *Gantry Gantry Gantry
- Area
- Nuclear Medicine
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare Nuclear Medicine issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare Nuclear Medicine issues
- How do I capture and print COR and Camera Uniformity (CUA) result screens on Genieacq, then send to a processing station for ACR
- Collision and E-stop lights are red on the back of the gantry and no movement with Hand Held Controller (HHC) or the genieacq.
- Need part numbers for the Millennium MG/MC table E-Stop
- How do I Delete a file on Genieacq that the system thinks is still acquiring
- VG, the contour works... but in the opposite way.
- MG_MC_MyoSight_MPR_MPS Bar Phantom Resolution Size