Pular para o conteúdo principal
GE Healthcare · CTLightSpeed18318033003

28139 Getting scanner hardware stopped scan with Error

Symptom

Scanning aborted with unexpected Loss Of High Voltage.

Error message

Error: 183180 I/O board fault detected. Relay K245 open, HVDC Bus not active. Cannot close backup contactor. EEBCSTAT = 1 KILLBC = 0 See Long Description for complete error Sun Jun 5 23:17:07 2005 Error: 183180 Host: OBC Process: Generator File: gen_rotor.c 1.39 Line: 450 Function: SYSTEM POWER CONTROL : HVDC Control Scan: 33003/1/1 Type: Scout Exception Level: Pri/Most Severe Time: 23:17:06:735 Log Series: 425 I/O board fault detected. Relay K245 open, HVDC Bus not active. Cannot close backup contactor. EEBCSTAT = 1 KILLBC = 0.

Applies to

Models
LightSpeed
Product
LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2208852 - HP INVERTER RHAPSODE - used on CT/i, 1.x, 2.x, 3.x and 4.x systems. 2208852 HP INVERTER RHAPSODE 2208852 HP INVERTER RHAPSODE Part #: 2208852 Part Type: GP REPAIRABLES Part Weight: 0.00 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: HP INVERTER RHAPSODE. FRU: Yes Status: Orderable Comments: IF RHAPSODE SYSTEM CURRENTLY HAS T3202JK CAN USE 2208852 . THE LONGER BOLTS NEEDED FOR THE 2208852 INVERTER ARE NORMALLY SUPPLIED IN A BAG WITH THE INVERTER OR CAN BE ORDERED SEPARATELY UNDER P/N 46-208561P43
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required