Symptom
Intermittent Scan Aborts at beginning of scan, Wrong angle Scouts (0 degree image looks more like 90 degree), All laser lights do not light. See LD for more details Intermittent Scan Aborts at beginning of scan, Wrong angle Scouts (0 degree image looks more like 90 degree), All laser lights do not light. 3 symptoms were seen 1 – Axial scans intermittently Abort at the beginning of the scan of the scan. 2 – Scouts scans intermittently Abort at the beginning of the scan. In addition sometimes if a 0 degree Scout is prescribed, when the picture is reconstructed it doesn’t look like it was taken at 0 degrees. It looks more like a Scout image taken at 90 degrees. 3- The laser lights intermittently do not all light. Specifically the external lights were mentioned as not being on. (or came on but because the gantry was not rotating to the correct location. The lights did not shine through the laser light cutouts in the gantry)
Error message
Error: 183946 Axial Encoder Feedback Error See Long Description for complete error Mon Oct 17 08:26:48 2005 Error: 183946 Suite: SMCT Host: STC Process: Axial File: ax_error.c Method: No Method Line: 1114 Function: AXIAL CONTROL : Axial Servo Control Loop Scan: 28848/2/1 Type: Helical Exception Level: Pri/Most Severe Time: 08:26:47:775 Log Series: 141 Axial Encoder Feedback Error Accumulated counts measured since last re-reference were 56005 different than expected. (tolerance = 30 counts) Probable cause: faulty axial encoder Possible cause: faulty axial control board. Error: 183921 Axial conditions at FAULT: See Long Description for complete error Mon Oct 17 08:26:48 2005 Error: 183921 Suite: SMCT Host: STC Process: Axial File: ax_error.c Method: No Method Line: 1114 Function: AXIAL CONTROL : Axial Servo Control Loop Scan: 28848/2/1 Type: Helical Exception Level: Pri/Most Severe Time: 08:26:47:775 Log Series: 141 Axial conditions at FAULT: velocity_fb = 2662 position_fb = 50389 Formatted data: 8100 100 8 7 1 0 1 1 101 1 0 2662 -12655275 13081394 2741 2582 134 133 0x019e38
Applies to
- Models
- HiSpeed
- Product
- HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-296854P2 Encoder - used on HSA/CTi/1.X gantries as the Axial Encoder and on Zeus Tables as the Elevation Encoder. See Long Description for part details Part #: 46-296854P2 Part Type: GP CONSUMABLES Part Weight: 0.88 LBR Repairable: No Returnable: No Hazardous Material: N Description: ELEVATION ENCODER FRU: Yes Status: Orderable ICV/Transfer Price: 104.55 USD /Each List Price: 815.40 USD /EACH Comments: FLS - ADDED PER PCN 189960. MS4636 4/22/97 & PCN 184587 MS03330 ** FE ASSIST ** Axial encoder HSA/CTi/1.X mounts on Axial Motor. (212001261 8/5/08) Part #: 46-296854P2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Axial Encoder FRU: Yes Status: Orderable Parts Offered as: Current: 46-296854P2 Used: Not Available Exchange: Not Available Comments: FLS - ADDED PER PCN 189960. MS4636 4/22/97 & PCN 184587 MS03330 5/5/97 COLLECTION OF DEFECTIVES START7/23/01 MS5277 FLS - PLEASE CONVERT SPH ORDER TO OVERNIGHT SHIPMENT IF PART AVAILABLE IN MEMPHIS AND IF FEDEX SERVICE (INCLUDING 8:00 AM SERVICE) MEETS FE NEED TIME CRITERIA FE ASSIST: IN ORDER TO INCREASE REVENUE GLOBAL PARTS HAS CREATED A PROGRAM TO HARVEST PARTS FROM RENEWABLE RESOURCES. THESE PARTS OFFER DIRECT SAVINGS TO THE BUSINESSAND THE FIELD. ORDER THE "U" PART IF AVAILABLE. IF NO "U" USED PARTS ARE AVAILABLE CARES WILL CHANGE THE ORDER TOA NEW PART.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.