Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT260134634284000225000

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Scan aborts with table cradle not moving during scan Scan aborts and table cradle not moving during scan with Error: 260134634 Tue Sep 27 16:15:22 2005 Error: 260134634 Host: tgp Process: tgp_subsys File: gtcb_cmc_profile.cxx Line: 537 Tue Sep 27 16:15:23 2005 Host : Table Sub-System Ermes # : 260134634 Exception Class : Secondary Severity : Pri/Soft File : gtcb_cmc_profile.cxx Line# : 537 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Cradle task received invalid corrected speed from cradle monitor task. received 284000 um/s expected 1 to 225000 um/s

Error message

Error: 260134634 Cradle task received invalid corrected speed from cradle monitor See LD for complete error May 12 06:36:02 2005 Error: 260134634 Host: tgp Process: tgp_subsys File: gtcb_cmc_profile.cxx Line: 537 Thu May 12 06:36:02 2005 Host : Table Sub-System Ermes # : 260134634 Exception Class : Secondary Severity : Pri/Soft File : gtcb_cmc_profile.cxx Line# : 537 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Cradle task received invalid corrected speed from cradle monitor task. received 284000 um/s expected 1 to 225000 um/s

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5245271 GTCB3 (Global Table Control Board 3) This board replaces 5142177-3 GTCB2 and GTCB part # 5112470 thru 5112470-5, and 5142177. See long description for details Part #: 5245271 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: GTCB3 BOARD ASSY FRU: Yes Status: Orderable Comments: added per to ECO2080741 by sso#305010201 in May 8, 2009. Part will be available end of Sep, 2009, please order old p/n before new p/n on shelf.
Area
CT

Full cause & step-by-step resolution