GE Healthcare · CTLightSpeed260112011
LightSpeed 2.X (Plus, QX/i) - Octane
Symptom
Table will not move up or down during a elevation table reference. Table will not move up or down during a elevation table reference
Error message
Error: 260112011 Received Table Rx but the Table was NOT READY Error : 260112011 Received Table Rx but the Table was NOT READY Thu Sep 1 08:49:43 2005 Error: 260112011 Host: tgp Process: tgp_subsys File: uif_table_tilt_mgr.c Line: 3197 Thu Sep 1 08:49:44 2005 Host : Table Gantry Processor Ermes # : 260112011 Exception Class : Soft Severity : Pri/Soft File : uif_table_tilt_mgr.c Line# : 3197 Function : No System Function Reported Scan Type : Helical Scan: 8439/4/1 Received Table Rx but the Table was NOT READY for Scanning.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-288170G1 ELEV/TILT AMP See LD for more 46-288170G1 ELEV/TILT AMP Part #: 46-288170G1 Part Type: GP CONSUMABLES Part Weight: 1.36 LBR Repairable: No Returnable: No Hazardous Material: N Description: ELEV/TILT AMP FRU: Yes Status: Orderable Comments: FLS - B: THE PROBLEM IS USUALLY WITHTHE 24 V RELAY. PLEASE HAVE THE FE TRY PART 2120791-9. MSA6521 5/29/01 DELISTED T3202CF 12/00 MS5277 FLS - PLEASE CONVERT SPH ORDER TO OVERNIGHT SHIPMENT IF PART AVAILABLE IN MEMPHIS AND IF FEDEX SERVICE (INCLUDING 8:00 AM SERVICE) MEETS FE NEED TIME CRITERIA FE ASSIST: IN ORDER TO INCREASE REVENUE GLOBAL PARTS HAS CREATED A PROGRAM TO HARVEST PARTS FROM RENEWABLE RESOURCES. THESE PARTS OFFER DIRECT SAVINGS TO THE BUSINESSAND THE FIELD
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required