Symptom
DAS converter board divide by zero and stage overflow during alpha calibrations
Error message
Error: 260007937 DAS Converter Board(s) encountered a stage overflow during alpha calibration. See Long Description for full error. Tue Dec 28 09:41:32 2004 Error: 260007937 Host: orp Process: orp_subsys File: brdflt.c Line: 82 Tue Dec 28 09:41:31 2004 Host : Das Ermes # : 260007937 Exception Class : Soft Severity : Sec/Soft File : brdflt.c Line# : 82 Function : No System Function Reported Scan Type : None/Unknown Scan: 50444/1/0 DAS Converter Board(s) encountered a stage overflow during alpha calibration. See Following Error Message for Converter Card List Board : 87 ASIC/Channel : 0x012c Stage : 1 Board Channel : 41 Error: 260007774 List of Converter Cards From the Previous Error Message Card List: xx xx See Long Description for full error Tue Dec 30 09:25:17 2008 Error: 260007774 Host: orp Process: orp_subsys File: CnvBoardInterface.cc Line: 852 Tue Dec 30 09:25:17 2008 Host : Das Ermes # : 260007774 Exception Class : Secondary Severity : Sec/Soft File : CnvBoardInterface.cc Line# : 852 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 List of Converter Cards From the Previous Error Message Card List: 8,11,14-15,19,21,25-27,29-36,38-58,60-63 Note on th
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2254888-2 TWIN 8 slice Elastomers BOX OF 25 ( DUROMETER ) 2254888-2 TWIN 8 slice Elastomers BOX OF 25 ( DUROMETER ) The Gdas uses the twin elastomer for both the 8 slice and 4 slice systems. only 16 slice systems use interposers 2254888-2 TWIN 8 slice Elastomers BOX OF 25 ( DUROMETER ) The Gdas uses the twin elastomer for both the 8 slice and 4 slice systems. only 16 slice systems use interposers Part #: 2254888-2 Part Type: GP CONSUMABLES Part Weight: 0.24 LBR Repairable: No Returnable: No Hazardous Material: N Description: 8 SLICE 70 DUROMETER FRU: Yes Status: Orderable Comments: FLS - B TWIN ELAST. (25PCS) IN CONTAINER IF YOU ORDER QTY ONE OF THIS PART YOU WILL GET BOX OF 25 MS13452 PER AVNET: THEY NO LONGER SUPPLY THIS ITEM. MS27911 DDI, CURRENT SUPPLIER FOR CTM MS13452
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.