GE Healthcare · CTLightSpeedVCT260132417260132499260132495
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Attention The detector is not at the proper temperature. see long description for more detail. Attention The detector is not at the proper temperature. If you wish to proceed with this scan select the "OK" button, however IQ may be degraded including image artifacts. Please contact GE Service
Error message
Error: 260132417 Detector Temperature Is High. Image quality may be affected. Current Reading:49.4901 Deg C A/D Counts:255 See LD for complete error Tue Sep 6 05:49:22 2005 Error: 260132417 Host: orp Process: orp_subsys File: ThermalManager.cxx Line: 1317 Tue Sep 6 05:49:21 2005 Host : Das Ermes # : 260132417 Exception Class : Secondary Severity : Sec/Soft File : ThermalManager.cxx Line# : 1317 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Detector Temperature Is High. Image quality may be affected. Current Reading:49.4901 Deg C A/D Counts:255 Error: 260132499 Rail Thermal sensor failure (shorted) Error code: 0x4 See Long Description for complete error. Tue Sep 6 05:49:22 2005 Error: 260132499 Host: orp Process: orp_subsys File: ThermalManager.cxx Line: 927 Tue Sep 6 05:49:20 2005 Host : Das Ermes # : 260132499 Exception Class : Secondary Severity : Sec/Soft File : ThermalManager.cxx Line# : 927 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Rail Thermal sensor failure (shorted) Error code: 0x4 Error: 260132495 Plenum Fan speed error Error code: 30 See LD for complete error Tue Sep 6 05:49:41 2005 Error: 260132495 Host: o
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5129997 VCT RAIL THERMISTOR SERVICE KIT Contains all 3 sensors and attachment tape for wiring See LD for details Part #: 5129997 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: VCT Rail Thermistor Service Kit FRU: Yes Status: Orderable Comments: Added 2/25/05 (212001261) VCT Top Critical Part is returnable, please return per RMA assigned when you order this part
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.