Pular para o conteúdo principal
GE Healthcare · MRSigna224152822415302241520

Will Not Scan with Gradient Errors.

Signa LX 9.1. The X axis minigram was shorted

Symptom

Will Not Scan with Gradient Errors.

Error message

2241528 % Axis Gradient Overload Occurred the GPM has reported an overload Fri Sep 10 10:58:56 2004 Error: 2241530 Host: wb001 Process: NSP File: HPC_PORT_ID:GAP/GIP/GP Line: 2540 Z Axis Gradient Overload Occurred. The Z Axis of the GPM has reported an overload. An overload is the result of the GPM output voltage not following the input command.

2241520 The GIP reported a X Axis Current Distortion Fault Tue May 16 17:26:51 2000 Error: 2241520 Host: MRCV_OW0 Process: NSP File: HPC_PORT_ID:GAP/GIP Line: 2231 The GIP reported a X Axis Current Distortion Fault. Another fault reported in the log could be the cause of this fault. Possible causes: GRAM Tuning, miniGRAM, components of the X Axis GPM, GIP, or cables.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
2169727-4 or T3503TT Mini GRAM Assembly
Area
MR