GE Healthcare · MRSigna2225462
Black Bands through anatomy Due to 8607 Gradient amp
Symptom
MRIQ: Black Bands through anatomy See the attached example image.
Error message
2225462 The Y axis Master Gradient amplifier has reported an overload.
Applies to
- Models
- Signa
- Product
- Signa Advantage Includes 4.x and 5.x systems. The software that this solution is tied to is indicated by the Software concept below.
- Software
- 5.X
- Component
- 46-287347P1 or T3406AC 8607 Gradient Amplifier
- Area
- MR
Full cause & step-by-step resolution