Pular para o conteúdo principal
GE Healthcare · CT26000100750042260001022260001012

Scanning aborted with DAS power errors

Symptom

"Scan hardware Stopped Scan" with DAS power errors. "Scan hardware Stopped Scan" with DAS power errors

Error message

Error: 260001007 CAN serial link error. A CC error was detected when sending a message to the DCB See LD for more details Wed Nov 22 05:44:35 2006 Error: 260001007 Host: OBC Process: HOP File: hop_cc_send.c Line: 196 Function: No System Function Reported Scan: 50042/6/0 Type: None/Unknown Exception Level: Pri/Soft Time: 05:44:35:070 Log Series: 80 CAN serial link error. An CC error was detected when sending a message to the DCB. Message code: 1004 Node ID: 51 Object ID: 0 Send status: 5

Error: 260001022 Communications between the DCB sub-system and OBC have been lost NOTE: Error: 260001022 can also state CCB to OBC comm problem but is a different error in PSDB! See the long description for the full error code Fri May 16 15:02:46 2003 Error: 260001022 Host: OBC Process: HOP File: hop_init_state.c Line: 249 Function: No System Function Reported Scan: 2476/200/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 15:02:47:210 Log Series: 501 Communications between the DCB sub-system and OBC have been lost. Attempting to establish communications.

Error: 260001012 The Rotating Controller Interface Bus (RCIB) fault line has been asserted See Long Description for the full error co

Applies to

Product
Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2218157 SDAS 5VD DIGITAL POWER SUPPLY WITH BRACKET - used on H1 gantry See LD for more 2218157 DAS 5VD DIGITAL POWER SUPPLY WITH BRACKET Part #: 2218157 Part Type: GP CONSUMABLES Part Weight: 3.64 LBR Repairable: No Returnable: No Hazardous Material: N Description: DAS 5VD DIGITAL PWR SUPPLY W/BRACKET FRU: Yes Status: Orderable Comments: FLS - DAS 5VD DIGITAL PWR SUPPLY W/BRACKET U/O LIGHTSPEED 3-23-99 MS08992 GWENN HQT: SHORT TERM SITUATION UNTIL 2218157 AVAILABLE FROM MFGING CAN USE 46-251198P3 THIS IS THE POWER SUPPLY WITHOUT THE BRACKET. 3/23/99 MS08992 PCN: 2218157 GELS - "DAS 5VD DIGITAL PWR SUPPLY W/B RACKET U/O LIGHTSPEED IN THE EVENT THAT IT IS NOT AVAILABLE, 46-251198P3 CAN BE USED (THIS IS WITHOUT THE BRACKET) SO THE FE WILL NEED TO HANG ON TO HIS OLD BR",POWER SUPPLY 5V
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.