GE Healthcare · CTLightSpeed260004000500402600010075004226000104950133
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane
Symptom
"Scan hardware Stopped Scan" with DAS power errors. "Scan hardware Stopped Scan" with DAS power errors
Error message
Error: 260004000 Collimator firmware object: Lost communications with controller GCAN error sending message = 5. See Long Description for Details. Error: 260004000 Host: OBC Process: Collimator File: ss_comm.c Line: 329 Function: SYSTEM COMMUNICATIONS : Rotating Controller I/F Bus Scan: 9553/3/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 14:37:12:150 Log Series: 377 Collimator firmware object: Lost communications with controller GCAN error sending message = 5. On different software version, the error may look like: Fri Feb 10 16:47:32 2006 Error: 260004000 Host: OBC Process: Collimator File: ss_comm.c Line: 329 023 Function: SYSTEM COMMUNICATIONS : Rotating Controller I/F Bus Scan: 50040/1/0 Type: None/Unknown Line: 374 Exception Level: Secondary Time: 16:47:32:525 Log Series: 491 Collimator firmware object: Lost communications with controller GCAN error sending message = 5. Error: 260001007 CAN serial link error. A CC error was detected when sending a message to the DCB See LD for more details Wed Nov 22 05:44:35 2006 Error: 260001007 Host: OBC Process: HOP File: hop_cc_send.c Line: 196 Function: No System Function Reported Scan: 50042/6/0 Type: None/Unknown Except
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-170021P75 15A 250 V DAS power fuse. See Long description for details P/N 46-170021P75 15A 250 V DAS power fuse Part #: 46-170021P75 Part Type: GP CONSUMABLES Part Weight: 0.01 LBR Repairable: No Returnable: No Hazardous Material: N Description: 015.000A 0250V 3AB SLOW UL FRU: Yes Status: Orderable
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.