Recon fails and calibrations fail to run

Symptom

Recon and calibrations fail to run See Long Description Recon fails and calibrations fail to run with Error: 230015005 Unable to retrieve the requested cal module from the Cal Database & Error: 230015021 Failure while writing data segment into cal file

Error message

Error: 230015021 Failure while writing data segment into cal file Sun Jun 19 20:37:53 2005 Error: 230015021 Host: wodct03403 Process: accs File: cal_db_api.c Line: 1639 Function: CAL Processing : SBC Processing Failure while writing data segment into cal file. File: /usr/g/cal_db/misc/zslope/4x250/zscal.0

Error: 230015005 Unable to retrieve the requested cal module from the Cal Database Sun Jun 19 21:05:13 2005 Error: 230015005 Host: wodct03403 Process: scanmgr File: cal_db_api.c Line: 451 Function: CAL Processing : SBC Processing Unable to retrieve the requested cal module from the Cal Database

Applies to

Models
HiSpeed, VCT
Product
HiSpeed QX/i - GRE Always MDAS 4 QX_P_M4 HiSpeed QX/i - GRE Method of visual identification Small rounded gantry similar in appearance to LS 3.X and 4.X. Always has an MDAS 4 slice Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. xxx
Software
04MW13.7 for all LightSpeed systems with GRE/Xtream consoles (QXi, PLUS, ULTRA, LS16, PRO 16, RT)(1.X, 2.X, 3.X, 4.X, 5.X) 04MW13.7.(gan)_(tube)_(DAS)_(recon) for all LightSpeed systems with GRE/Xtream consoles (1.X, 2.X, 3.X, 4.X, 5.X) This is the new common Apps software for GRE/Xtream consoles. The GRE Common Applications software supports multiple hardware configurations via a single software set. The new software version string has the following format: <software_label>.<G>_<T>_<D>_<R> The "software label" format is: YYLWxx.ZZ Where: YY = Year. For example 04. L = Location. M - Milwaukee, H - Hino, B - Beijing. Wxx = Fiscal Week. For example, W15. ZZ = Revision number for each build/reroll. Refer to Table B for an explanation of the <G>_<T>_<D>_<R> format. Table B Common Applications Sotware Naming Convention G = GANTRY TYPE T = TUBE TYPE D = DAS TYPE R = RECON HW TYPE H1 = LightSpeed 1.X P = Performix S4 = SDAS4 P = Pegasus H2 = LightSpeed 2.X, 3.X, 4.X O = OEM M4 = MDAS4 G = GRE HP = LightSpeed 5.X H = Hercules M8 = MDAS8 WB = LightSpeed 5.x RT M16 = MDAS16 QX = HighSpeed QXi G16 = GDAS16 The variables G, T, D, and R are generated dynamically based on the type of hardware configuration this software is installed on. The information for the dynamic part of the string is obtained from the INFO files which are located in /usr/g/config directory. An example string based on the above format is as follows: 04MW13.7.H2_P_M4_G The break down is as follows: This is 2004 software release at Milwaukee in Fiscal Week 13 with a revision number "7". The hardware configuration is "H2" gantry with "Performix" tube, "MDAS4", and "GRE" recon hardware. CDL Link: Refer to: DIRECTION 2378261-100 GRE SOFTWARE INSTALLATION PROCEDURES
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.