PROTEUS Tomo-Link Generator Selection for Proteus systems
Symptom
PROTEUS The Tomol-Lnk documentation is not explicit regarding the type of generator to select when there is a Proteus ( Jedi ) used. The incorrect sweep times will be displayed on the console if the incorrect generator type is selected. 8 degree settings will be OK 20 degree slow = 1 second, should be 2 seconds 20 degree fast setting will be OK 40 degree slow = .5 second, should be 4 seconds 40 degree fast = 1 second, should be 2 seconds
Applies to
- Product
- PROTEUS XRa
- Component
- Tomolink JEDI Generator This generator is used on Proteus and Precision 500D systems.
- Area
- X-ray R&F
Full cause & step-by-step resolution