Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT260132511260132756260132768

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Customer can not scan with DAS temperature errors. Customer can not scan with DAS temperature errors

Error message

Error: 260132511 Plenum filter clogging. Service may be required soon. See LD for complete error Mon Jun 6 04:09:00 2005 Error: 260132511 Host: orp Process: orp_subsys File: ThermalManager.cxx Line: 1192 Mon Jun 6 04:08:59 2005 Host : Das Ermes # : 260132511 Exception Class : Secondary Severity : Sec/Soft File : ThermalManager.cxx Line# : 1192 Function : No System Function Reported Scan Type : Axial Scan: 1784/2/3 Plenum filter clogging. Service may be required soon.

Error: 260132756 Digital Interface Board(s) Detected An Invalid Hardware Tracker Table on a Digital Module See Long Description for complete error Fri Jul 29 06:30:09 2005 Error: 260132756 Host: orp Process: orp_subsys File: Ifb.cxx Line: 435 Fri Jul 29 06:30:08 2005 Host : Das Ermes # : 260132756 Exception Class : Secondary Severity : Pri/Soft File : Ifb.cxx Line# : 435 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Digital Interface Board(s) Detected An Invalid Hardware Tracker Table on a Digital Module See Following Error Message for Digital Interface Board List DIFB : 1 A/D Board: 0 Thu Aug 4 12:20:55 2005 Error: 260132756 Host: orp Process: orp_subsys File: Ifb.cxx Line: 435 Thu Aug 4

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues