IPG Sequencer RAM walking 1s and 0s test failed Due to the IPG II Board

Symptom

Intermittent Image Reconstruction Failures The system intermittently fails to reconstruct images. The failure occurs after the acquisition has completed.

Error message

2222267 IPG Sequencer RAM walking 1s and 0s test failed. see long description Error seen with TPS diags: 1094647477 1 1 Wed Sep 8 12:44:37 2004 2222267 2 lx-mr diags diagsMain.cc 2263 IPG BOARD LEVEL DIAGS FAILURE IPG Sequencer RAM walking 1s and 0s test failed. Pattern Written: 00000000H Pattern Expected: 00000000H Pattern Read: 00000080H Address: 00d80002H EN 291 SR 310

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution