GE Healthcare · X-ray R&F
AMX4 will not drive in a forward direction due to a worn motor brush.
Symptom
AMX4 will not drive, operator getting message release handle and errors 214 and 217.
Replaced the motor brushes FE must order two for each motor. See attached parts list.
Error message
Left & Right Drive Stall AMX4 Errors 13, 14 214, 217
Applies to
- Product
- AMX4 AMX4+
- Area
- X-ray R&F
Full cause & step-by-step resolution