GE Healthcare · MRSigna2236568223670522367071718752236394
Intermittent Ghosting or Smearing of Images
Symptom
MRIQ: Intermittent Ghosting or Smearing of Images See attached Files. Intermittent ghosting on images. Not specific to any type of scan. No errors are seen in the log when scanning. TPS Diags had failures. Shown below.
Error message
2236568 CERD BOOT ERROR MESSAGE - ON CAP - ASC DIAGS ERROR Fri Feb 25 14:26:22 2000 Error: 2236568 Host: lx-mr Process: NSP File: TPS:prgasc.c Line: 281 CERD BOOT ERROR MESSAGE - ON CAP - ASC DIAGS ERROR Failed while running asc power up diags. Status: ffffffffH 2236705 Cerd Rcvr Diags failure - Analog rcv Anti Alias Filter Test Failure Cerd rcv diags fail Proc diags Error 2236705 Analog rcv anti alias filter test failure Rcv channel X has poor freq response. Cerd rcv diags fail Error 2236705 Analog rcv anti alias filter test failure Rcv channel X has poor freq response. 2236707 CERD Diags : Analog Receiver Narrowband Filter Two Tone Test Failure Fri Feb 25 14:30:56 2000 Error: 2236707 Host: lx-mr Process: diags File: diagsMain.m Line: 2110 CERD Diags : Analog Receiver Narrowband Filter Two Tone Test Failure Receiver channel 1 has large third-order intermodulation products. FFT Mag: 5713, Min.Allowed: 0, Max.Allowed: 1132, Freq.Hz: 171875 2236394 CERD DIAGS - AIME Functional Test Failure. Fri Feb 25 14:26:22 2000 Error: 2236394 Host: lx-mr Process: NSP File: TPS:aimed_func.c Line: 372 CERD DIAGS - AIME Functional Test Failure. Output point 3 of a 256 point FFT is incorrect. The
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 8.2.5
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.