GE Healthcare · MRSigna
SPT Head Quick Check fails SNR with values in all three planes of about 23
Symptom
SPT Head Quick Check fails SNR with values in all three planes of about 23. The mobile driver reports that Head SNR is about 23 in all three planes in SPT Head Quick Check. He reseated the coil connections, no help. I asked the driver to reboot the scanner and reseat the coil connections again, this time on both sides of the I & Q coax cables and the Quick-Disconnect box, and rerun the test. He did this but it didn't fix the problem.
Error message
A serious problem has been detected which will impact image quality. This message appears to the operator at the end of the SPT Head Quick Check test: "A serious problem has been detected which will impact image quality".
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- 2148308-2 HEAD COIL QUICK DISCONNECT (NORMAL POLARITY) 1.5T
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.