Pular para o conteúdo principal
GE Healthcare · MRSigna222533422253372224733223216622254212225412

Clarify #007627 SPI MDS Link Failures

Symptom

Scan Aborts with MDS Link Error

Error message

2225334 SRI failed to respond to a data packet from SPI

2225337 SPI-SRI Communication Error. SRI failed to respond to a watchdog packet from SPI. Error: 2225337 Host: MRS2 OC0 Process: NSP File: HPC PORT ID:spi sri watchdog fail Line: 13 SPI-SRI Communication Error. SRI failed to respond to a watchdog packet from SPI. Watchdog packets waiting for a response: 0x0001

2224733 - The SRI is not responding. Please reset the TPS chassis and try again.

2232166 Error during TPS reset. Fri Sep 26 07:45:32 1997 Suite: MR01 Host: OC0 Proc: spectro Error: 2232166 File: SpAcqCtrl.m 1.3 Method: tpsAlive: Line: 472 Error during TPS reset. Please try again and press CANCEL upon successful reset

2225421 SPI MDS link failure. The SPI received an MDS broke error packet from the SCB. Fri Sep 26 07:45:41 1997 Suite: MR01 Host: OC0 Proc: HPC_PORT_ID Error: 2225421 File: PMH.m 1.29 Method: spi_mds Line: 1000 SPI MDS link failure. The SPI received an MDS broke error packet from the SCB. The link appears to be broken between the SPC & SCB. Check rf cabinet power, the SPC board, & the connections & signal level of the fiber optic link between SPC & SCB. Fri Sep 26 07:45:41 1997 Suite: MR01 Host: OC0 Proc

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.