GE Healthcare · MRSigna22446452244597

LX Communication problem between GP and GSC.

Symptom

MRIQ Scan stops Multiple Times, They Restart the Scan and it Completes but Images Have Severe Zipper Artifacts. http://3.87.118.28/optec3/common/8xdocs/epst/mod11/m11_010_000.html Link to image quality troubleshooting guide. Scan stops in the middle of gradient stressing (GRE) sequences. After the gradients recovered, the scan completes (eventually after several stops). The images get zippers and artifacts due to the stops.

Error message

2244645 - The GSC Stopped Responding to the Gradient Processor (GP). Thu Dec 11 09:31:21 2003 2244645 2 MR-OC0 NSP SCP:GP 471 The GSC stopped responding to the Gradient Processor (GP). Verify that the GSC is powered up and the serial communication cable is attached. If this does not work, try cycling GSC power.

2244597 - Gradients recovered successfully on all 3 axis. There are no Faults at the Present Time. Thu Dec 11 09:31:26 2003 2244597 2 MR-OC0 NSP SCP:GP 1069 The gradient system on the Z axis recovered successfully. There are no faults at the present time.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.0
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.