Configuration File is Corrupt

Symptom

Configuration File is Corrupt One of the config files is corrupt. For an example, see the Long Description. For example, in the Coil Config file (/w/config/CoilConfig.cfg) we see this entry for the HFO Head coil: coil1 = { startFastRec = "4" \ ATTN_I = "0" \ quadRotation = "0" \ " \ reconScale = "1.78 <---- note the "\ is on the wrong side of this line quadRcvCoil = "0" \ headDefaultFreqDir = "1" \ mcErrorEnable = "3" \ cableLoss = "1.129" \ mcReconEnable = "0" \ numFastRec = "0" \ ATTN_Q = "0" \ autoshimRcvr = "0" \ mcPortEnable = "3" \ extremity = "no" \ cfoption = "31" \ xmitAtten = "0" \ coilName = "HEAD" \ coilCode = "MRIDHEADPA" \ coilType = "1" \ fastTGstartRG = "12" \ fastTGstartTA = "150" \ startRec = "0" \ numRec = "4" \ endRec = "3" \ multiCoil = "yes" \ mcChannelSel = "15" \ linearQuad = "1" \ endFastRec = "4" \ coilLoss = "0.0224" }

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR