Symptom

"Scan hardware Stopped Scan" with PDU errors "Scan hardware Stopped Scan" with PDU errorss

Error message

Error: 183949 Could NOT CLOSE Gantry (Axial) Loop Contactor See LD for full error CT01 Host: STC Proc: Axial Error: 183949 Function: AXIAL CONTROL : Axial Loop Contactor Interlocks Gantry (Axial) LOOP CONTACTOR Error. Could NOT CLOSE Gantry (Axial) Loop Contactor Another example: Tue May 24 08:41:21 2005 Suite: CT02 Host: STC Proc Axial Error: 183949 File: ax_error.c Method No Method 1114 Function: AXIAL CONTROL : Axial Loop Contactor Interlocks Scan: 65001/11/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 08:41:21:940 Log Series: 60 Gantry (Axial) LOOP CONTACTOR Error. Could NOT CLOSE Gantry (Axial) Loop Contactor. Axial Gantry Command/Status:0x9e28. EN 401 SR 442

Error: 183182 I/O board pilot relays are closed. PDU pilot relay not energized. HVDC Rail Voltage is xxxx See Long Description for more details. Error: 183182 I/O Bd. pilot relays are closed. PDU pilot relay not energized. HVDC is xxxx. ** Note the HVDC Rail Voltage. It is important in troubleshooting Tue Sep 21 06:13:59 2004 Error: 183182 Host: OBC Process: Generator Function: SYSTEM POWER CONTROL: HVDC Control Scan: 50001/1/1 Type: Axial Exception Level: Pri/Most Severe I/O board pilot relays are closed. 

Applies to

Models
LightSpeed, BrightSpeed
Product
LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.