GE Healthcare · MRSigna5017365043

MRIQ Shear Distortion Oval Images on DWI Diffusion Weighted Images.

Clarify 025959 LX 9.1.

Symptom

MRIQ Shear Distortion Oval Images on DWI Diffusion Weighted Images. See long description. Some of the diffusion-weighted epi images are reconstructed as ovals tipped at a -45 degree angle. This is a 1.5T Hi-Speed LX system. All EPT cals pass. SPT passes. No other imaging mode has this effect. A System Health check shows no problem with the system configuration or calibrations. All REQUIRED (but not all) eddy curent crossterms are applied. If at the OnLine Center, you may view a phantom image of this problem: system ID 610973MR, Exam 50173, Series 1, Image 121. Normal phantom images are: same exam, same series, images 69 and 101. Note that the oval phantom image looks exactly like the example shown in Service Note 65043 "Diffusion Weighted Image Artifacts" page 4 "Shear Distortion due to eddy currents on logical read x diffusion", second image from the right.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution