Pular para o conteúdo principal
GE Healthcare · MRExciteOvation225008122185372250080

CWA ID19545. Scan Stops in Middle of Scans.

Symptom

Scan Stops in Middle of Scans See Long Description Hard failure during scanning, APS works ok, AP recon error watching APS window.

Error message

2250081 Z Axis Gradient Data Overrange Occurred. The GP detected an error in the incoming High Speed Data from the STIF/SRF. Wed Jun 29 06:40:35 2005 Error: 2250081 Host: lx-mr2 Process: NSP File: SCP:GP Line: 1981 Z Axis Gradient Data Overrange Occurred.^A The GP detected an error in the incoming High Speed Data from the STIF/SRF. Swap fiber optics at the Systems Cabinet. Run fiber optic diagnostics.

2218537 Scan stopped, Got Recon-Scan Stop packet With Error Message Key Fri Apr 16 10:11:34 2004 Error: 2218537 Host: LX1 Process: scn File: MsaScanMgr.m Line: 2641 Scan stopped, Got Recon-Scan Stop packet with error message key - 2248

2250080 Y Axis Gradient Data Overrange Occurred. The GP detected an error in the incoming High Speed Data from the STIF/SRF. Wed Jun 29 05:36:19 2005 Error: 2250080 Host: lx-mr2 Process: NSP File: SCP:GP Line: 1981 Y Axis Gradient Data Overrange Occurred.^A The GP detected an error in the incoming High Speed Data from the STIF/SRF. Swap fiber optics at the Systems Cabinet. Run fiber optic diagnostics.

Applies to

Models
Excite, Ovation
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.