Unable to perform Filament Aging Calibration after a LFC

Symptom

Unable to perform Filament Aging Calibration after a LFC Unable to perform Filament Aging Calibration (other generator faults possible) after performing a Software Load from Cold for any reason or replacing the Generator Power Module.

Error message

36000130 Generator Abort Error, 8400006 Generator error retry For Optima systems look for error code 36000130 Generator Abort Error, Debug text = Djinn error 0805H: System Identifier mismatch For Brivo systems look for error code 8400006 Generator error. Retry. If persists call service, Root cause = 0x805ABORT System Identifier mismatch

Applies to

Models
Optima
Product
Optima XR220amx Optima XR200amx
Software
Software Versions for Filament Aging Issue on Portable Units
Area
X-ray R&F