Next button in the lower right corner of the 2nd page of the Edit View window is grayed out.
Symptom
Next button in the lower right corner of the 2nd page of the Edit View window is grayed out. See Long Description for Details. In the Protocol Editor (Left Monitor Utilities > Preferences > Protocol Database > Edit) , when editing a view (Chest_1_View LAT, for example) that uses the Tethered Detector (Digital Cassette) and has horizontal flip applied by default ("Apply horizontal flip to image" box checked), the Next button in the lower right corner of the 2nd page of the Edit View window will be grayed out (see the attached picture). This doesn't allow the change made in the previous screen (unchecking the "Apply horizontal flip to image" box) to be saved.
Applies to
- Models
- Discovery, Optima
- Product
- Discovery XR650 Optima XR640
- Software
- XR650 and XR640 M3 See Long Description for Config.txt file. atlasidc-2.1-6 atlas_linux_release-7.9-0 calypso-8.70-2 can_pci266_2615_smp-7.9-ctt627 CSD-2.4A-1 detector-1.1-6 dmcots-1.1-15 dm_Platform_release-1.1-15 ea3-3.2-5 enggui-15.4-3 FlightRecorder-15.13-ATL.7.9 GlobalOneChinaFW_V4-1-2 IIPClassA-3.3.12-3 IIPClassM-3.3.12-3 IPCots-4.2-1 iui_globalone_release-5.1-23 j2re-1_4_2_13-linux-i586 j2sdk-1_4_2_13-linux-i586 jedi_lightning_magic-6.33BF6-1 machineDB-2-5 magic-aux_tools-1-1 magic-config_prod_GlobalOne-1-1 magic-cots-4-58 mcf_release-1.7-0 pnf-2.5-3 pos_download_mke-1-22 psys_can_pci266_2615_smp-7.9-ctt627 rcim2_smp_linux-7.9-0 RCOC-1.2.8-0 RTIP_G1_Magic_Release-FW17-RC3 saber_release-2.3.64C-1 SysConfig_Linux-103B-ATL.7.8.6 tao_linux_release-7.9-0 CD-R Label Data: 1) Business Name: Global Diagnostic X-Ray 2) CD-R Part Number: 5316167-3 3) Revision: 4 4) Software Title: Application Software 5) Disc Number: 1 of 1 6) Product Name: Discovery XR650/Optima XR640 7) Legally Required Messages ONLY: 8) Duplication Date: 9) Copyright Date: 2009
- Area
- X-ray R&F
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray R&F issues
- Collimator light does not work from the RCIM2 even though the button is present
- Discovery XR650 or Optima XR640
- No procedure to copy IP looks from a Definium 8000 system to a Discovery XR650 or Optima 640 system
- When the ./xrdinstall.om script is run to install the Operator Manual and a specific language is selected (English, for example), nothing happens.
- Grid (any focal length) insertion is not recognized by the wallstand.
- Detector calibration BadPixel issue