Pular para o conteúdo principal
GE Healthcare · MRSigna2252305

2252305 TR Driver Fault due to Faulty Head Heliax Cable Termination.

Symptom

Cannot scan Head coil, error 2252305 TR driver fault detected - open circuit Source of fault is Head T/R on Connector A Value 370 mA, limit 500 mA

Error message

2252305 TR driver fault detected - open circuit Source of fault is Head T/R on Connector A Value 370 mA, limit 500 mA

Applies to

Models
Signa
Product
Signa HDxT Apps. For use with HD16.0_VO1 and HD16.0_VO2 version software only. For use with HD16.0_VO1 and HD16.0_VO2 version software only.
Software
HD16 Includes HD16.0_VO1 and HD16.0_VO2 version software and applies to 1.5T and 3.0T "Signa HDxT Apps" systems only. Includes HD16.0_VO1 and HD16.0_VO2 version software and applies to 1.5T and 3.0T "Signa HDxT Apps" systems only.
Area
MR

Full cause & step-by-step resolution