GE Healthcare · MRDiscovery22522892224666
Unable to scan due to Driver Module in Service Mode.
Symptom
Unable To Scan With Driver Module Errors Unable To Scan With Driver Module Errors.
Error message
2252289 Driver Module is in Service Mode. A Cause is DD Fault Disable Switch. Error: 2252289 Host: MR3T Process: NSP File: SCP:DCB Line: 933 Driver Module is in Service Mode.^A Cause is DD Fault Disable Switch.-> 2224666 Auto Prescan failed, Scanning has failed because the driver module is in service mode and the patient ID is not 'geservice' See Long Description Error: 2224666 Host: WMHS3TMR Process: psc File: ../PscMainControl.cc Line: 1383 Exception Class: Unknown Function: Prescan: Host Autoprescan Auto Prescan failed, Scanning has failed because the driver module is in service mode and the patient ID is not 'geservice'.-> 2252289 Driver Module is in Service Mode. A Cause is TR Fault Disable Switch. Error: 2252289 Host: MR3T Process: NSP File: SCP:DCB Line: 933 Driver Module is in Service Mode.^A Cause is TR Fault Disable Switch.->
Applies to
- Models
- Discovery
- Product
- Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
- Software
- DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA ID100513. Will Not Scan with Gradient Errors.
- Multicoil Bias Fault errors on multiple coils.
- Corrupt Image Database.
- Auto Prescan Failures and RF amp setup incomplete Errors Due to Defective RF amp.
- The system is unable to scan with any surface coil, get dynamic disable faults.
- CWA ID103384. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.