Pular para o conteúdo principal
GE Healthcare · MRSigna2249439116092249431116808

2249439 The VxWorks System Integrity Checker could not be run on all loaded modules due to defective DRF2 card in MGD ch…

2249439 The VxWorks System Integrity Checker could not be run on all loaded modules due to defective DRF2 card in MGD chassis.

Symptom

TPS reset fails with 2249439 The VxWorks System Integrity Checker could not be run on all loaded modules.

Error message

2249439 The VxWorks System Integrity Checker could not be run on all loaded modules see long Error: 2249439 Host: mr1 Process: diagCoord File: sysIntegrity.c Line: 1076 Task ID: 0x7fffb60 Time: 11609 Object: APS Exception Class: Unknown Function: Software/Firmware/Config: N/A The VxWorks System Integrity Checker could not be run on all loaded modules. semTake failed for semaphore ID 0x7FFFDC0. Errno = 0x3D0004. This may happen if a module takes too long to load.

2249431 A PCI error was detected, but message queue send failed to the exception logging task see long description Error: 2249431 Host: GEMR_OC4 Process: diagCoord File: mgd820PciErr.c Line: 1378 Task ID: 0x3e2b6e0 Time: 116808 Object: APS Exception Class: Unknown Function: Software/Firmware/Config: N/A A PCI error was detected, but message queue send failed to the exception logging task; errno = 0x3D0002. Harrier Err Enable Reg: 0x00000000; Harrier Err Status Reg: 0x00000000. Harrier Err Address Reg: 0x00000000; Harrier Err Attribute Reg: 0x8000.

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
2298332 DIGITAL RECEIVER AND FILTER II DRF2 2298332 DIGITAL RECEIVER AND FILTER II DRF2 Part #: 2298332Part Type: GP CONSUMABLES Part Weight: -3175168.25 LBRRepairable: NoReturnable: NoHazardous Material: NDescription: DIGITAL RECEIVER AND FILTER II DRF2FRU: YesStatus: OrderableICV/Transfer Price: 858.16 USD /Each ICV/Transfer Price: 858.16 USD /Each List Price: 8939.61 USD /EACH
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues